DATASCOPE CORP. - MAHWAH CS100 SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number N/A
Device Problem Battery Problem (2885)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/28/2020
Event Type  malfunction  
Manufacturer Narrative
A getinge field service engineer (fse) was dispatched to investigate. The fse evaluated the iabp unit and observed that the iabp unit was not working on battery mode. The fse checked the power supply battery charging voltage and observed that the voltages were within factory specifications. The fse suspected that the battery may be defective; however, decided to let the batteries charge for 2 days as a troubleshooting measure as the batteries may have been deeply discharged. A supplemental report will be submitted when additional information is provided.
 
Event Description
It was reported that the charging indicator for the cs100 intra-aortic balloon pump (iabp) was not showing on the display. It is unknown under which circumstances the event occurred or if a patient was involved; however there was no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCS100
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
Manufacturer Contact
dorota wolpiuk
1300 macarthur blvd
mahwah, NJ 07430
MDR Report Key11059835
MDR Text Key225339816
Report Number2249723-2020-02193
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K031636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial,Followup,Followup,Followup
Report Date 05/05/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/22/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberN/A
Device Catalogue Number0998-00-3013-55
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received04/24/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/01/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-