RESMED LTD ASTRAL 150 DBL KIT FRA Back to Search Results
Model Number 27074
Device Problems Unintended Ejection (1234); Failure to Run on Battery (1466); Battery Problem (2885); Protective Measures Problem (3015)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/20/2020
Event Type  malfunction  
Manufacturer Narrative
The device was returned to resmed and an evaluation confirmed the complaint.The internal battery and pcb were replaced to address the issues.The device was serviced and fully tested before it was returned to the customer.(b)(4).
 
Event Description
It was reported to resmed that an astral device displayed a battery error message and the pcb had capacitor leakage.There was no patient harm or serious injury reported as a result of this incident.
 
Manufacturer Narrative
The astral device was returned to resmed for an investigation.Review of the device data logs confirmed the reported battery error message and revealed an internal battery degraded warning alarm and an error message (sf180) related to a battery charger fault.Based on all available evidence and complaint investigations of a similar nature, the investigation determined that the battery error message, sf180 and internal battery degraded warning alarm were due to an isolated component failure within the device battery assembly.Resmed's risk analysis for this failure mode concludes that the risk is acceptable.Resmed reference#: pr 2167255.
 
Event Description
It was reported to resmed that an astral device displayed a battery error message and the main circuit board had a leaky super capacitor.There was no patient harm or serious injury reported as a result of this incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASTRAL 150 DBL KIT FRA
Manufacturer (Section D)
RESMED LTD
1 elizabeth macarthur drive.
bella vista
sydney, nsw 2153
AU  2153
MDR Report Key11062055
MDR Text Key224598000
Report Number3004604967-2020-01252
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
PMA/PMN Number
K152068
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Type of Report Initial,Followup
Report Date 02/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/23/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number27074
Device Catalogue Number27074
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/29/2020
Distributor Facility Aware Date01/17/2021
Date Manufacturer Received01/17/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-