Model Number 16-02-85
Device Problem Pumping Problem (3016)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/01/2020
Event Type  Malfunction  
Manufacturer Narrative

There was no patient involvement. Livanova (b)(4) manufactures the heater-cooler system 3t. The incident occurred in (b)(6). A livanova field service technician was dispatched to the customer facility. The error code on the patient pump was confirmed. In order to solve the issue, the distribution board and the complete patient bridge were replaced. The unit was tested and put back in service in its expected function. Based on the above, it cannot be ruled out that the issue on the patient pump was caused by the defective motor electronic inside the motor of the replaced pumps. This led to the defect on the safety resistors for the pump on the distribution board.

 
Event Description

Livanova (b)(4) received a report that a heater-cooler system 3t displayed an error message associated to a patient pump during the procedure. There was no report of patient injury.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameHEATER-COOLER SYSTEM 3T
Type of DeviceCONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80309
GM 80309
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80309
GM 80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada 80004
MDR Report Key11064966
MDR Text Key223765594
Report Number9611109-2020-00711
Device Sequence Number1
Product Code DWC
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberK191402
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type HEALTH PROFESSIONAL
Reporter Occupation OTHER HEALTH CARE PROFESSIONAL
Remedial Action Other
Type of Report Initial
Report Date 12/18/2020
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received12/23/2020
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL Number16-02-85
Was Device Available For Evaluation? No Answer Provided
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received12/01/2020
Was Device Evaluated By Manufacturer? Yes
Date Device Manufactured04/14/2018
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No