Catalog Number 0998-00-3023-53 |
Device Problem
Premature Discharge of Battery (1057)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 11/30/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge service territory manager (stm) performing preventative maintenance (pm) replaced the batteries to resolve the issue.The stm completed all safety, functionality and calibration checks and all tests passed to factory specifications.The iabp unit was cleared for clinical use and released to the customer.(b)(6).
|
|
Event Description
|
It was reported that during preventive maintenance (pm) performed by a getinge service territory manager (stm), the cs300 intra-aortic balloon pump (iabp) failed the battery run time test.There was no patient involvement, and no adverse event reported.
|
|
Event Description
|
It was reported that during preventive maintenance (pm) performed by a getinge service territory manager (stm), the cs300 intra-aortic balloon pump (iabp) failed the battery run time test.There was no patient involvement, and no adverse event reported.
|
|