• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NXSTAGE MEDICAL, INC. NXSTAGE SYSTEM ONE; HIGH PERMEABILITY HEMODIALYSIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NXSTAGE MEDICAL, INC. NXSTAGE SYSTEM ONE; HIGH PERMEABILITY HEMODIALYSIS SYSTEM Back to Search Results
Model Number CAR-500
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/07/2020
Event Type  malfunction  
Manufacturer Narrative
The cartridge was discarded and was not available for evaluation.A device history record (dhr) review was conducted for the reported lot number and revealed the product was released for distribution having met quality and manufacturing specifications and requirements.The instructions for use states "make sure mated luer-connectors are secure but do not over-tighten, especially when connections are wet".
 
Event Description
A report was received on (b)(6) 2020 from the nurse of a critical care patient with unspecified pathology, who stated the cartridge venous line luer broke during a continuous renal replacement therapy (crrt) treatment on (b)(6) 2020, there was no indication of adverse impact to patient and no report of medical intervention being required.
 
Manufacturer Narrative
Analysis of this event by quality engineering confirmed that the product design prevents any risk to patient safety should the connector failure recur as reported by the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NXSTAGE SYSTEM ONE
Type of Device
HIGH PERMEABILITY HEMODIALYSIS SYSTEM
Manufacturer (Section D)
NXSTAGE MEDICAL, INC.
350 merrimack street
lawrence MA 01843
MDR Report Key11067202
MDR Text Key231785962
Report Number3003464075-2020-00086
Device Sequence Number1
Product Code KDI
UDI-Device IdentifierM535CAR5000
UDI-Public+M535CAR5000/$$062200878008V
Combination Product (y/n)Y
PMA/PMN Number
K133547
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 02/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/23/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/01/2022
Device Model NumberCAR-500
Device Catalogue NumberCAR-500
Device Lot Number00878008
Was Device Available for Evaluation? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-