• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number N/A
Device Problem Battery Problem (2885)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
A getinge field service engineer (fse) was dispatched to evaluate the unit.The fse verified the capacity of one battery was empty, tried to swap the battery to other slot but still would not charge.To fix the issue, the fse replaced the battery pack and performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.
 
Event Description
It was reported that prior to use, the cardiosave intra-aortic balloon pump (iabp) had a battery pack that would not charge.There was no patient involvement, and no adverse event reported.
 
Event Description
It was reported that prior to use, the cardiosave intra-aortic balloon pump (iabp) had a battery pack that would not charge.There was no patient involvement, and no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
MDR Report Key11068961
MDR Text Key224599516
Report Number2249723-2020-02223
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108421
UDI-Public10607567108421
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 01/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/23/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number0998-00-0800-45
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Date Manufacturer Received12/22/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-