SMITH & NEPHEW, INC. MOTOR DRIVE UNIT HAND CNTRL PWRMX EL; SAW, POWERED, AND ACCESSORIES
|
Back to Search Results |
|
Model Number 72200616 |
Device Problem
Protective Measures Problem (3015)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/13/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed on the product and observed no issue.A functional evaluation revealed a hand piece sensor fault.The complaint was confirmed and the root cause has been associated with an electrical component failure.Factors that could have contributed to the reported event include a failure of the reed switch or damage on the hall board which may contribute to a short circuit.Internal complaint reference: (b)(4).
|
|
Event Description
|
It was reported that the mdu was not working.No case involved.Preliminary results of investigation have concluded the mdu revealed a hand piece sensor fault; therefore, makes it a reportable event.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
Manufacturer Narrative
|
H6 h10: the reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed on the product and observed no issue.A functional evaluation revealed a hand piece sensor fault.The complaint was confirmed and the root cause has been associated with an electrical component failure.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.Factors that could have contributed to the reported event include a failure of the reed switch or damage on the hall board which may contribute to a short circuit.
|
|
Search Alerts/Recalls
|
|
|