Device Problem
Application Program Freezes, Becomes Nonfunctional (4031)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/09/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
|
|
Event Description
|
A customer reported that the system froze before surgery.The scheduled procedure was cancelled.
|
|
Manufacturer Narrative
|
The company service representative examined the system, confirmed and replicated the reported event of a system freeze.The nonconforming touchscreen was replaced to address the issue.The system was then tested and met all product specifications.The system manufacturing device history record (dhr) was reviewed.Based on qa assessment, the product met specifications at the time of release.The root cause of the reported event can be attributed to a nonconforming touchscreen.Data will continue to be monitored for evidence of adverse trending with further action taken, as appropriate.The manufacturer internal reference number is: (b)(4).
|
|