HEARTSINE TECHNOLOGIES LTD PACKAGE, RED 350P, 1X PAD-PAK-03, GERMAN, 350-STR-ST-10; AUTOMATED EXTERNAL DEFIBRILLATOR
|
Back to Search Results |
|
Model Number SAM 350P |
Device Problems
Device Alarm System (1012); Noise, Audible (3273)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/11/2020 |
Event Type
malfunction
|
Event Description
|
Audible alarm, intermittent red led.No error displayed in the memory.No patient involvement.
|
|
Manufacturer Narrative
|
Red status led and beep confirmed upon receipt.Upon investigation the device indicated no failed self-tests.Measurements taken indicated a fault with mosfet q45, which forms part of the red status indicator logic circuitry, confirmed by replacement of the component and clearance of the fault.
|
|
Event Description
|
Audible alarm, intermittent red led.No error displayed in the memory.No patient involvement.
|
|
Search Alerts/Recalls
|
|
|