Evaluation of the returned device did not confirm the reported issue of the mixer being out of tolerance by 10%.
The fio2 knob was found to be bent but did not affect the functionality of the device.
The mixer functioned as designed.
At this time there is no evidence that a manufacturing non-conformity contributed to the reported complaint, and the instructions for use were reviewed and determined to provide adequate instructions and warnings for the safe and effective use of the device.
Therefore, no corrective or preventative actions are necessary.
All complaints are trended and reviewed by management on a monthly basis.
As part of this monthly review, any excursion above the control limits for this failure mode will be assessed, documented and acted upon as warranted.
Manufacturer reference file no.
(b)(4).
|