Catalog Number 106697 |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 12/07/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that during priming with a prismaflex m100 set, an external fluid leak was observed at the access pod of the dialyzer.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
H10: a batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The device was received for evaluation.A pressure test was performed and a leak was observed from the filter pod.A visual inspection observed the pod's membrane was pierced resulting in the observed leak.The reported condition was verified.The cause is supplier.Related.Should additional relevant information become available, a supplemental report will be submitted.
|
|