THORATEC CORPORATION HEARTMATE 3 LVAS IMPLANT KIT VENTRICULAR (ASSISST) BYPASS Back to Search Results
Model Number 106524US
Device Problem Infusion or Flow Problem (2964)
Patient Problems Atrial Flutter (1730); High Blood Pressure/ Hypertension (1908); Right Ventricular Dysfunction (2054); Heart Failure/Congestive Heart Failure (4446)
Event Date 12/15/2020
Event Type  Injury  
Manufacturer Narrative
No further information was provided. A supplemental report will be submitted when the manufacturer¿s investigation is completed.
 
Event Description
It was reported that the patient's device had low flows and low voltage advisories due to normal battery depletion. A computed tomography angiography did not show occlusion. The patient is hypertensive and has right ventricular dysfunction with new flutter/rapid ventricular response - likely cause of low flows. The low flows were resolved through blood pressure and heart rate improvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameHEARTMATE 3 LVAS IMPLANT KIT
Type of DeviceVENTRICULAR (ASSISST) BYPASS
Manufacturer (Section D)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer (Section G)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer Contact
bob fryc
6035 stoneridge drive
pleasanton, CA 94588
7818528204
MDR Report Key11089782
MDR Text Key224393963
Report Number2916596-2020-06344
Device Sequence Number1
Product Code DSQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P160054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date05/23/2022
Device Model Number106524US
Device Catalogue Number106524US
Device Lot Number7093825
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received01/15/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/18/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 12/29/2020 Patient Sequence Number: 1
-
-