BLUE BELT TECHNOLOGIES ANSPACH EMAX 2 PLUS HAND PIECE - ROHS; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number PFSR101209
Device Problem Overheating of Device (1437)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/03/2020
Event Type  malfunction  
Event Description
It was reported that, during a navio preventive maintenance, the anspach drill overheated and has a tear in the tubing.No patient was involved.No other complications were reported.
 
Manufacturer Narrative
The anspach drill intended for use in treatment was returned for evaluation.The reported problem was visually confirmed.The jacket is torn 1/8"" from the end of the drill.Although the reported problem was visually confirmed, a functional evaluation was performed on the part beyond the reported complaint.The evaluation followed pv0005 rev.B.The evaluation failed.High pitched sound and grinding coming from the drill.Drill runs hot to the touch as well.A review of manufacturing and service records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints associated with the part with the failure modes ¿overheating of device | material split, cut or torn¿ and ¿damaged cord, overheating¿ identified similar events.The most likely cause of the tear in the outer sheath improper handling.Care and caution should be exercised during the surgical site setup and tear down to protect the device cable and from sharp objects or from situations that pin the cable between two objects.Do not use excessive force on the device strain reliefs during the decontamination process to minimize separating the cable from the strain relief.Refer to the navio surgical system user's manual and the navio surgical system instrument kit cleaning and sterilization guide for proper handling.The most likely cause of the overheating is mechanical failure of the drill motor shaft.The drill is an oem part and cannot be further disassembled to determine a root cause.Based on the investigation, no further containment or corrective action is recommended or required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANSPACH EMAX 2 PLUS HAND PIECE - ROHS
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
MDR Report Key11089854
MDR Text Key224589331
Report Number3010266064-2020-02133
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
PMA/PMN Number
K191223
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 02/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberPFSR101209
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/21/2020
Date Manufacturer Received01/28/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-