MAUDE Adverse Event Report: CAREFUSION SD 8300 ALARIS ETCO2 MODULE PUMP, INFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION SD 8300 ALARIS ETCO2 MODULE PUMP, INFUSION Back to Search Results
Model Number 8300
Device Problems Break (1069); Crack (1135); Failure to Calibrate (2440); Appropriate Term/Code Not Available (3191)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
The customer reported problem was confirmed. The device was repaired, retested, and released back to the customer. A review of the device history record in sap for sn (b)(4) was performed from the date of the manufacture to date of the release of product, which confirmed that this device was not involved in a production failure, and product was returned for servicing which correlates to the customer reported issue. A review of the complaint history record in the trackwise was performed for the sn (b)(4) which confirmed no similar complaints with the same or related failure mode. The customer stated that there was no patient involvement.
 
Event Description
Failed preventive maintenance iui- isolation rib damage oridion board- failed co2 cal test damaged/cracked - case front, housing alarm - failure to alarm- (b)(4). "no disposable connected".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name8300 ALARIS ETCO2 MODULE
Type of DevicePUMP, INFUSION
Manufacturer (Section D)
CAREFUSION SD
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer (Section G)
CAREFUSION SD
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer Contact
kalaiarasan govindan
10020 pacific mesa blvd
san diego, CA 92121-4386
8586172000
MDR Report Key11094199
MDR Text Key233725237
Report Number2016493-2020-77854
Device Sequence Number1
Product Code CCK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K031741
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 10/25/2019
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/30/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number8300
Device Catalogue Number8300
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/15/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-