MAUDE Adverse Event Report: CORDIS CORPORATION OPTEASE RETRIEVAL FILTER FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORDIS CORPORATION OPTEASE RETRIEVAL FILTER FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Catalog Number 466F220A
Device Problem Unintended Movement (3026)
Patient Problems Coagulation Disorder (1779); Internal Organ Perforation (1987); Perforation (2001); Perforation of Vessels (2135); Obstruction/Occlusion (2422); Thrombosis/Thrombus (4440); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 08/10/2000
Event Type  Injury  
Manufacturer Narrative
It was reported that a patient underwent placement of an optease vena cava filter. The information provided indicated that the filter subsequently malfunctioned and caused perforation, tilting, caval thrombosis, deep vein thrombosis (dvt) and device unable to be retrieved with no documented attempts to remove the filter. The indication for the filter implant, procedural details and medical history of the patient have not been provided and there is currently no additional information available for review. The product was not returned for analysis and the sterile lot number has not been provided; therefore, no device analysis nor device history record review could be performed. The optease vena cava filter is indicated for use in the prevention of recurrent pulmonary embolism (pe) via percutaneous placement in the vena cava for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pe where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pe where anticoagulant therapy has failed, or is contraindicated. The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart. Blood clots, dvt and occlusive thrombosis within the filter and vasculature do not represent a device malfunction. Clinical factors that may have influenced the event include patient, pharmacological and vessel characteristics. Ivc filters are not indicated for use in the prevention of dvt. Ivc filter tilt has been associated with operator technique and vessel anatomy, specifically asymmetry and tortuosity. The instructions for use notes vessel damage such as intimal tears and perforation as procedural and long-term complications related to ivc filters. The timing and mechanism of the tilt and perforation has not been reported at this time and a clinical conclusion could not be determined as to the cause of the event. Without images available for review the reported events could not be confirmed or further clarified. Given the limited information available for review at this time, there is nothing to suggest that the reported events are related to the design and manufacturing process of the device; therefore, no corrective action will be taken. Should additional information become available, the file will be updated accordingly.
 
Event Description
As reported in the legal brief, a patient underwent placement of an optease vena cava filter. The filter subsequently malfunctioned and caused injury and damage to the patient, including, but not limited to perforation, tilting, caval thrombosis, deep vein thrombosis (dvt) and device unable to be retrieved with no documented attempts to remove the filter. As a direct and proximate result, the patient suffered life-threatening injuries and damages and required extensive medical care and treatment. As a further proximate result, the patient has suffered and will suffer significant medical expenses, pain and suffering and other damages.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameOPTEASE RETRIEVAL FILTER
Type of DeviceFILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
CORDIS CORPORATION
14201 nw 60th avenue
miami lakes FL 33014 2802
Manufacturer (Section G)
CORDIS CORPORATION
14201 nw 60th ave
miami lakes FL 33014 2802
Manufacturer Contact
karla castro
14201 nw 60th ave
miami lakes, FL 33014-2802
7863138372
MDR Report Key11094981
MDR Text Key224361636
Report Number1016427-2020-04679
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K034050
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial,Followup
Report Date 02/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date09/30/2016
Device Catalogue Number466F220A
Device Lot Number16004155
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/22/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/13/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 12/30/2020 Patient Sequence Number: 1
-
-