• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CS300 SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CS300 SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Catalog Number 0998-00-3023-53
Device Problems Use of Device Problem (1670); Appropriate Term/Code Not Available (3191)
Patient Problem No Patient Involvement (2645)
Event Date 12/08/2020
Event Type  malfunction  
Manufacturer Narrative
A getinge service territory manager (stm) was dispatched to evaluate the iabp unit. The stm was able to reproduce the issue when the iabp was turned on. The coiled cable was not fully seated. The stm re-seated the coiled cable then restarted the iabp. The error was no longer there. The stm completed all safety, functionality and calibration checks and all tests passed to factory specifications. The iabp unit was cleared for clinical use and released to the customer. (b)(6).
 
Event Description
It was reported that the cs300 intra-aortic balloon pump (iabp) had an electrical test error code 120. There was no patient involvement, and no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCS300
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
Manufacturer Contact
dorota wolpiuk
1300 macarthur blvd
mahwah, NJ 07430
MDR Report Key11096705
MDR Text Key226017928
Report Number2249723-2020-02251
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation
Type of Report Initial,Followup
Report Date 02/10/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/30/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number0998-00-3023-53
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/14/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-