Catalog Number 0998-00-0800-53
Device Problem Charging Problem (2892)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Date 12/08/2020
Event Type  malfunction  
Manufacturer Narrative
A getinge service territory manager (stm) was dispatched to evaluate the iabp unit. The stm determined the iabp was not seated in the cart properly. The stm realigned the cart and console until it seated properly and the batteries charged. A full pm was performed and the batteries charged overnight to a full charge. The stm completed all safety, functionality and calibration checks and all tests passed to factory specifications. The iabp unit was cleared for clinical use and released to the customer. (b)(6).
 
Event Description
It was reported that the cardiosave intra-aortic balloon pump (iabp) would not charge and the iabp would not work off the wall charge. It is unknown the circumstances under which the event occurred. It is also unknown if there was patient involvement. However, there was no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCARDIOSAVE HYBRID DOMESTIC
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
Manufacturer Contact
dorota wolpiuk
1300 macarthur blvd
mahwah, NJ 07430
MDR Report Key11096945
MDR Text Key226042229
Report Number2249723-2020-02254
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation
Type of Report Initial,Followup
Report Date 01/29/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/30/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number0998-00-0800-53
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/04/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-