• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SCREW CAPTURE SLEEVE; PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SCREW CAPTURE SLEEVE; PLATE, FIXATION, BONE Back to Search Results
Model Number 71177345
Device Problem Corroded (1131)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/10/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that, the screw capture sleeve was found rust.Incident occured during inspection.No case reported; therefore, there was no patient involvement.
 
Manufacturer Narrative
G4, h2 and h6: the device, used in treatment, was returned for evaluation.A visual inspection of the returned device confirmed the stated failure mode.The device was found to be rusty, rendering the device inoperable.The device was manufactured in 2019 and shows signs of extensive use.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCREW CAPTURE SLEEVE
Type of Device
PLATE, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11099950
MDR Text Key224525481
Report Number1020279-2020-08102
Device Sequence Number1
Product Code HRS
UDI-Device Identifier03596010601728
UDI-Public03596010601728
Combination Product (y/n)N
PMA/PMN Number
K123055
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 04/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/31/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71177345
Device Catalogue Number71177345
Device Lot Number19LM08125
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/29/2020
Date Manufacturer Received03/28/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-