• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - FAIRFIELD LINEAR 7.5 FR. 34CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - FAIRFIELD LINEAR 7.5 FR. 34CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0684-00-0479-07
Device Problem Leak/Splash (1354)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/09/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(6).The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.(b)(4).
 
Event Description
It was reported that during insertion of the intra-aortic balloon (iab), there was a backflow of blood noticed in the tubing.There was no patient harm or adverse event reported.
 
Manufacturer Narrative
The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.The failure mode is addressed in the risk file and is operating within its risk profile.The ifu addresses the reported failure.There were no ncmrs identified which could cause or contribute to the reported failure.The investigation does not indicate that the device was inadvertently released as non-conforming or an adulterated product or was a counterfeit.The complaint history review did not identify an adverse trend (increase in number of complaints over past three (3) months).Based on the rational provided above, no escalation to the capa process is required.Reference complaint #: (b)(4).H3 other text : device not returned.
 
Event Description
It was reported that during insertion of the intra-aortic balloon (iab), there was a backflow of blood noticed in the tubing.There was no patient harm or adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LINEAR 7.5 FR. 34CC IAB
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ 07004
MDR Report Key11100261
MDR Text Key224540863
Report Number2248146-2020-00694
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567109695
UDI-Public10607567109695
Combination Product (y/n)N
PMA/PMN Number
K041281
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/31/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/31/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/19/2022
Device Model Number0684-00-0479-07
Device Catalogue Number0684-00-0474
Device Lot Number3000104732
Was Device Available for Evaluation? Yes
Date Manufacturer Received03/31/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age80 YR
Patient Weight65
-
-