MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC EMAX 2 PLUS MOTOR; MOTOR, SURGICAL INSTRUMENT, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC EMAX 2 PLUS MOTOR; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Model Number EMAX2PLUS
Device Problems Noise, Audible (3273); Complete Loss of Power (4015); Excessive Heating (4030)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/09/2020
Event Type  malfunction  
Manufacturer Narrative
Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the reported condition was confirmed.The assignable root cause was determined to be due to component failure from normal wear.Udi: (b)(4).
 
Event Description
It was reported by (b)(6) that during service and evaluation, it was determined that the power of the motor device was insufficient/low, and the device generated heat and excessive noise.It was further observed that the flex circuit and cord were damaged, and the device had a component damage.It was further determined that the device failed pretest for visual assessment, cable assessment, motor thermistor assessment, rotational speed assessments, noise assessment and handpiece temperature assessment.It was noted in the service order that the device stopped working at 64,000 rotations per minute (rpms) and it made an unusual sound/noise.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMAX 2 PLUS MOTOR
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
kara ditty-bovard
4500 riverside drive
palm beach gardens, FL 33410
6103142063
MDR Report Key11100285
MDR Text Key224527808
Report Number1045834-2020-02205
Device Sequence Number1
Product Code GEY
UDI-Device Identifier00845384002055
UDI-Public00845384002055
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K080802
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other
Type of Report Initial
Report Date 11/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/31/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEMAX2PLUS
Device Catalogue NumberEMAX2PLUS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2020
Date Manufacturer Received12/23/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/15/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-