Model Number N/A
Device Problems Battery Problem (2885); Charging Problem (2892)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
A getinge service territory manager (stm) was dispatched to evaluate the iabp but the customer contacted the stm by phone to fix the issue. Customer reported that the unit was in rescue mode and not completely seated. Customer reseated unit and batteries started to charge. The iabp unit was back in use.
 
Event Description
It was reported that during a routine check, the battery of the cardiosave intra-aortic balloon pump (iabp) would not charge. There was no patient involvement, and no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCARDIOSAVE HYBRID DOMESTIC
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
Manufacturer Contact
dorota wolpiuk
1300 macarthur blvd
mahwah, NJ 07430
MDR Report Key11103131
MDR Text Key226042550
Report Number2249723-2020-02274
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 01/27/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/31/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberN/A
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/26/2021
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/11/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-