DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number N/A |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 12/10/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
During a preventative maintenance (pm) service, the getinge field service engineer (fse) discovered that iabp unit failed the drive manifold test.To fix the issue, the fse installed a new drive manifold, and performed drive manifolds test.The iabp unit passed all tests and pre-use checks.The fse then completed a full calibration, as well as all functional and safety checks to meet factory specifications.The iabp was then released to the customer for return to clinical service.(b)(6).
|
|
Event Description
|
It was reported that during a preventative maintenance (pm) service repair performed by a getinge field service engineer (fse), the cardiosave intra-aortic balloon pump (iabp) failed the drive manifold test.There was no patient involvement, and no adverse event reported.
|
|
Event Description
|
It was reported that during a preventative maintenance (pm) service repair performed by a getinge field service engineer (fse), the cardiosave intra-aortic balloon pump (iabp) failed the drive manifold test.There was no patient involvement, and no adverse event reported.
|
|