Model Number 0998-00-3023-55
Device Problem Battery Problem (2885)
Patient Problems No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/11/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(6). A getinge field service engineer (fse) evaluated the iabp unit and the unit failed the battery test. The fse replaced the ion batteries and the 9v doppler battery, and performed a pm with full calibration, functional and safety checks to meet factory specifications. Unit passed all calibration, functional and safety test per factory specifications. The iabp was then released to the customer and cleared for clinical service.
 
Event Description
It was reported that during preventive maintenance (pm) performed by a getinge field service engineer (fse), the cs300 intra-aortic balloon pump (iabp) failed the battery test. Battery supported only 87 min. There was no patient involvement, and no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCS300
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
Manufacturer Contact
dorota wolpiuk
1300 macarthur blvd
mahwah, NJ 07430
MDR Report Key11103214
MDR Text Key226018420
Report Number2249723-2020-02280
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeSN
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial,Followup
Report Date 06/30/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/31/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-3023-55
Device Catalogue Number0998-00-3023-55
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received06/30/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-