MAUDE Adverse Event Report: MICRO THERAPEUTICS, INC. DBA EV3 REACT CATHETER CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICRO THERAPEUTICS, INC. DBA EV3 REACT CATHETER CATHETER, PERCUTANEOUS Back to Search Results
Model Number REACT-71
Device Problem Device Damaged by Another Device (2915)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/10/2020
Event Type  Malfunction  
Manufacturer Narrative

If information is provided in the future, a supplemental report will be issued.

 
Event Description

Medtronic received a report that the react catheter had a protruding braid wire as found during a competitor investigation of an embotrap device and its interaction with the catheter. It was stated that the protruding braid wire prevented the withdrawal of the embotrap device into/through the catheter lumen, causing damage. The embotrap was removed with a stent retriever. No further additional information was available. The patient was undergoing thrombus removal for ais.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameREACT CATHETER
Type of DeviceCATHETER, PERCUTANEOUS
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer Contact
tricha miles
9775 toledo way
irvine, CA 92618
7635140379
MDR Report Key11103229
MDR Text Key225565021
Report Number2029214-2020-01374
Device Sequence Number1
Product Code DQY
Combination Product (Y/N)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type OTHER
Reporter Occupation
Type of Report Initial
Report Date 12/31/2020
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received12/31/2020
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL NumberREACT-71
Device Catalogue NumberREACT-71
Device LOT NumberUNKNOWN
Was Device Available For Evaluation? No
Is The Reporter A Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received12/01/2020
Was Device Evaluated By Manufacturer? No Answer Provided
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-