MAUDE Adverse Event Report: CAREFUSION SD 8100 ALARIS PUMP MODULE; PUMP, INFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION SD 8100 ALARIS PUMP MODULE; PUMP, INFUSION Back to Search Results
Model Number 8100
Device Problems Failure to Read Input Signal (1581); Device Sensing Problem (2917)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
This reported event and subsequent repairs were investigated through the service repair process.Failure data and parts-used information were reviewed for the sap and trackwise files and found relevant to the service repair.A review of the device service history record was performed from the date of manufacture to the date corresponding to this service notification number.The database showed quality notifications were opened for the device without correlation to the reported complaint.A review of the device history record in sap for sn (b)(4) was performed from the date of the manufacture to date of the release of product, which confirmed that this device was involved in a production failure, and product was returned for servicing which correlates to the customer reported issue.A review of the complaint history record in the trackwise was performed for the sn (b)(4) which confirmed no similar complaints with the same or related failure mode.The customer stated that there was no patient involvement.
 
Event Description
Failed preventive maintenance.Keypad- unresponsive key.Pressure sensor- patient side.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
8100 ALARIS PUMP MODULE
Type of Device
PUMP, INFUSION
Manufacturer (Section D)
CAREFUSION SD
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer (Section G)
CAREFUSION SD
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer Contact
prasanth maduraiveeran
10020 pacific mesa blvd
san diego, CA 92121-4386
8586172000
MDR Report Key11103509
MDR Text Key233447457
Report Number2016493-2020-86172
Device Sequence Number1
Product Code FRN
UDI-Device Identifier10885403810015
UDI-Public10885403810015
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133532
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 06/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/31/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number8100
Device Catalogue Number8100
Was Device Available for Evaluation? Yes
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/28/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-