Model Number 406849 |
Device Problem
Fracture (1260)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/04/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.Further information was requested but not received.
|
|
Event Description
|
During an atrial fibrillation ablation procedure, a crack was noted on the tip of the inner sheath following insertion into the patient (and blood can leak out of the sheath).The sheath was replaced with the same model of the device in order to complete the procedure with no adverse consequences to the patient.
|
|
Manufacturer Narrative
|
One 8.5f fast-cath introducer sheath and dilator were received for evaluation.Visual inspection revealed the dilator distal tip had been split.Functional testing revealed a leak at the hemostasis valve.The cap was removed from the hemostasis hub and the hemostasis seal was microscopically inspected.Tearing, resulting in a hole was noted in the seal.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the torn seal and subsequent leak is consistent with damage during use.The ifu states: do not remove dilator or catheter rapidly.Damage to the valve may occur, potentially compromising hemostasis.
|
|
Search Alerts/Recalls
|