Model Number N/A
Device Problems Gas Leak (2946); Output below Specifications (3004); Appropriate Term/Code Not Available (3191)
Patient Problem No Patient Involvement (2645)
Event Date 12/08/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(6). A getinge field service engineer (fse) evaluated the iabp unit and was able to reproduce the reported issue. The fse replaced the o-ring from the quick disconnect fitting and the helium bottle. The fse performed all functional and safety checks to meet factory specifications. Unit passed all functional and safety test per factory specifications. The iabp was then released to the customer and cleared for clinical service.
 
Event Description
It was reported that during a clinical training, the cardiosave intra-aortic balloon pump (iabp) gave a low helium alarm. There was no patient involvement, and no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCARDIOSAVE HYBRID TYPE E/F PLUG
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
Manufacturer Contact
dorota wolpiuk
1300 macarthur blvd
mahwah, NJ 07430
MDR Report Key11112574
MDR Text Key226045638
Report Number2249723-2020-02256
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup
Report Date 02/12/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received01/04/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberN/A
Device Catalogue Number0998-00-0800-55
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received02/12/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/23/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-