Model Number CLK-4
Device Problems Break (1069); Failure to Power Up (1476)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned to omsc but was returned to olympus repair center in hamburg for evaluation. Olympus repair center inspected the device and confirmed the following; the reported phenomenon was reproduced. The front panel was broken. There were small holes in the hose of the air supply line. The halogen lamp had low power and was dark. The air pump was difficult to operate under pressure. The bottom plate, screws, and rear panel were corroded by moisture. The reported event that the power switch failed appeared to be caused by aging deterioration because more than 12 years have passed since the device was manufactured. Olympus concluded that the device did not work because the device did not turn on due to a power switch failure. Device history record (dhr) review indicates that the product was manufactured and tested in accordance with all applicable procedures and met all final product release criteria. If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus medical systems corp. (omsc) was informed by the user that the power switch failed and the device did not work. There was no report of patient injury associated with the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameHALOGEN LIGHT SOURCE
Type of DeviceHALOGEN LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA 192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key11129040
MDR Text Key225598743
Report Number8010047-2021-00985
Device Sequence Number1
Product Code FAJ
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K993041
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation
Type of Report Initial
Report Date 01/07/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received01/07/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberCLK-4
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional?
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received12/09/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/09/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-