The device has not been returned to omsc but was returned to olympus repair center in hamburg for evaluation.
Olympus repair center inspected the device and confirmed the following; the reported phenomenon was reproduced.
The front panel was broken.
There were small holes in the hose of the air supply line.
The halogen lamp had low power and was dark.
The air pump was difficult to operate under pressure.
The bottom plate, screws, and rear panel were corroded by moisture.
The reported event that the power switch failed appeared to be caused by aging deterioration because more than 12 years have passed since the device was manufactured.
Olympus concluded that the device did not work because the device did not turn on due to a power switch failure.
Device history record (dhr) review indicates that the product was manufactured and tested in accordance with all applicable procedures and met all final product release criteria.
If additional information becomes available, this report will be supplemented.
|