MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH CATHETER, STEERABLE Back to Search Results
Model Number 4FC12
Device Problem Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/04/2021
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a cryoablation procedure, when the balloon catheter was inserted into the sheath, the physician could not prevent air from stopping when aspirating. A hemostatic valve issue was suspected. The sheath was replaced, but when the balloon catheter was inserted, a stream of air bubbles was noted. It was noted the balloon catheter was a tight fit going into the sheath. The balloon catheter was replaced, resolving the issue. The case was completed with cryo. No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameFLEXCATH ADVANCE STEERABLE SHEATH
Type of DeviceCATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA H9R 5Z8
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key11132134
MDR Text Key225538460
Report Number3002648230-2021-00017
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial,Followup
Report Date 06/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/07/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot Number0010396177
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/02/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received06/09/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/05/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 01/07/2021 Patient Sequence Number: 1
Treatment
2AF84 BALLOON CATHETER
-
-