• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. JGRKNT 1.0MM MINI 3-0 NDLS; FASTENER, FIXATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. JGRKNT 1.0MM MINI 3-0 NDLS; FASTENER, FIXATION Back to Search Results
Model Number 912082
Device Problem Fracture (1260)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/08/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Udi# (b)(4).Concomitant medical products: 912082 jgrknt 1.0mm mini 3-0 ndls 440000 912082 jgrknt 1.0mm mini 3-0 ndls 554590.Foreign-(b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034 - 2020 - 04456 0001825034 - 2020 - 04457.
 
Event Description
It was reported that during the surgery, the sutures of the juggerknot broke when the surgeon pulled the sutures after he inserted the anchors into the burr hole.There was a surgical delay of 15 to 30 minutes for preparing an alternate device.
 
Event Description
No additional event information is available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, d9, g3, g6, h1, h2, h3, h6, h10 complaint sample was evaluated and the reported event was confirmed.Visual examination of the returned products identified no damages to the main portion of the products (handles).Inspection confirms the returned sutures have been torn/broken.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JGRKNT 1.0MM MINI 3-0 NDLS
Type of Device
FASTENER, FIXATION
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key11139134
MDR Text Key225739018
Report Number0001825034-2020-04458
Device Sequence Number1
Product Code MBI
UDI-Device Identifier00880304523418
UDI-Public(01)00880304523418(17)250225(10)537470
Combination Product (y/n)N
PMA/PMN Number
K110879
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 02/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/08/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number912082
Device Catalogue Number912082
Device Lot Number537470
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/27/2021
Was the Report Sent to FDA? No
Date Manufacturer Received02/17/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-