Model Number N/A |
Device Problem
Fracture (1260)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/15/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported the device broke while impacting the femoral component.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.Visual inspection of the returned part confirm that the device was fractured and not all pieces were returned.The device also exhibited wear consistent with usage.Device history record (dhr) was reviewed and no discrepancies were found.The reported device has been in the field for approximately ten years and eleven months.It is unknown for how many times the device is being used.The root cause can be contributed to normal wear and tear of device from repeated use over time.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|