Model Number 288022
Device Problem Mechanics Altered (2984)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate. Udi: (b)(4). Investigation summary: the device was received and evaluated at the service center. The reported complaint that there was motor failure with the device was confirmed. It was found that the motor was sticking as it was rusty. The motor was replaced and the device was repaired, tested and found to be fully functional. Fluid ingress into the system and contact with the motor is responsible for the corrosion of the motor. The corroded motor has a tendency to stick and not turn, hence is responsible for the issue reported by the customer. The service history has been reviewed in lieu of the device history record for this device since it was previously serviced. The device was last serviced on 01/30/2020 and passed all functional testing before being returned to the customer. At this point in time, no corrective action is required, and no further action is warranted. Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Event Description
It was reported by the affiliate in germany that during service evaluation, it was determined that the handpiece device had an unspecified malfunction. During in-house engineering evaluation, it was determined that the motor was sticking as it was rusty there was no procedure nor patient involvement reported. No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name8022 HANDPC TORNADO SHAVER
Type of DeviceLINE-POWERED ARTHROSCOPIC SHAVER SYSTEM
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
MEDOS INTERNATIONAL SARL
chemin blanc 38
le locle CH-24 00
SZ CH-2400
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6013142063
MDR Report Key11139882
MDR Text Key225807553
Report Number1221934-2021-00051
Device Sequence Number1
Product Code HRX
Combination Product (y/n)N
PMA/PMN Number
K954465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial
Report Date 01/08/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received01/08/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number288022
Device Catalogue Number288022
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/20/2020
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/08/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-