Catalog Number 107144 |
Device Problems
Break (1069); Fluid/Blood Leak (1250)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that during the installation of prismaflex tpe2000 set, an external fluid leak was observed due to a break in a weld.There was no patient involvement reported.No additional information is available.
|
|
Manufacturer Narrative
|
H10: the actual sample was received for investigation.The visual inspection of the filter observed a disconnection from the support plate by the dialysate port.The reported condition was verified.The cause was likely due to a mechanical shock during transportation.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|