CONCORD MANUFACTURING 2008T HD SYS. CDX W/BIBAG BLUE STAR; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Model Number 191126
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Patient Involvement (2645)
Event Date 09/17/2020
Event Type  malfunction  
Manufacturer Narrative
Plant investigation: the functional board was returned to the manufacturer for physical evaluation.During the visual inspection, physical damage was identified on the board.The returned part exhibited evidence of burn damage.The functional board was installed onto a test machine with a known good ic20.The test machine powered on and functioned as intended.However, the crit-line verification failed.Upon further investigation, a shorted component was identified.The shorted component was a damaged inductor l16 and damaged tracing.A review of the device manufacturing records was conducted by the manufacturer.There were no deviations or non-conformances during the manufacturing process.In addition, a device history record (dhr) review was performed and verified that the results of the in-progress and final quality control (qc) testing met all requirements.Upon completion of the evaluation, the reported issue was confirmed.
 
Event Description
The biomedical technician (biomed) from a hemodialysis (hd) user facility reported that a crit-line device was not working on a fresenius 2008t hd machine.The biomed resolved the reported issue by replacing the functional board.There was no patient involvement associated with the reported issue.The functional board was returned to the manufacturer for evaluation.Upon physical evaluation of the returned part, evidence of burn damage was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2008T HD SYS. CDX W/BIBAG BLUE STAR
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer (Section G)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer Contact
matthew amaral
920 winter st
waltham, MA 02451
7816999758
MDR Report Key11141230
MDR Text Key225817319
Report Number2937457-2021-00048
Device Sequence Number1
Product Code KDI
UDI-Device Identifier00840861102099
UDI-Public00840861102099
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K173972
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 01/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/08/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number191126
Device Catalogue Number191126
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/25/2020
Is the Reporter a Health Professional? Yes
Device AgeMO
Date Manufacturer Received12/17/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/22/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-