Model Number TB-0535FCS
Device Problem Device Fell (4014)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/11/2020
Event Type  malfunction  
Manufacturer Narrative
The subject device has not yet been received for evaluation.The root cause cannot be determined at this time.If additional information becomes available, a supplemental report will be filed.
 
Event Description
It was reported that the tb-0535fcs probe broke and fell into the patient.The probe broke when users were creating a bladder flap on a laparoscopic hysterectomy.The broken probe was recovered with no patient injury.No user injury reported.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device was not returned to olympus for evaluation; therefore, the reported phenomenon or condition of the device could not be confirmed.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely the broken probe occurred due to contact with a surgical instrument or the non-insulated area of grasping section.The instructions for use have the following statement which can prevent the event: "do not activate output in seal & cut mode while the grasping section is closed without contacting tissue or vessel, or ensuring that tissue is transected.Otherwise, a local increase of the temperature due to a friction between the probe tip and the grasping section may result in various forms of damage in the probe tip and/or the tissue pad, such as premature wear, breakage, deformation, and/or falling off inside the body cavity and/or partial separating." "when cutting and vessel sealing is performed in seal & cut mode, apply light tension on the tissue so that users can confirm it is transected.Also, stop activation immediately after tissue is transected.Otherwise, the grasping section, the tissue pad, or the probe tip may break and fall off, and partial separating of the tissue pad may occur due to a local increase of temperature caused by the friction between tissue pad and the probe tip during activation." "the thunderbeat instrument should be used for soft tissue.Do not activate output while grasping hard tissue such as bone or highly calcified tissue, or hard objects such as metal clips, stapler, or other instruments (e.G., uterine manipulator, forceps, and others).Otherwise, it may cause the probe tip to be scratched or come into direct contact with the metal area of the grasping section as the heat generated by the friction between the hard object and the probe tip could cause wear/ deforming/ splitting/ protruding/ partial separating of the tissue pad.In turn, the probe may break before displaying an error window or generating an alarm tone." "do not grasp or let the probe tip contact hard objects such as metal clips, stapler, or other instruments (e.G., uterine manipulator).Also, be careful to avoid contacting the probe tip with those accidentally.Particularly during activation, a scratch on the probe tip could occur due to ultrasonic vibration, which leads the probe tip to break and fall off into the body cavity.In addition, the high-frequency (rf bipolar) current flows through the metal and generates spark discharge, which may cause burns and decrease functionalities.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THUNDERBEAT 5 MM, 35 CM, FRONT-ACTUATED GRIP TYPE S
Type of Device
ULTRASONIC SURGICAL DEVICE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key11141426
MDR Text Key225831759
Report Number8010047-2021-01129
Device Sequence Number1
Product Code GEI
UDI-Device Identifier04953170383540
UDI-Public04953170383540
Combination Product (y/n)N
PMA/PMN Number
K172610
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 03/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/08/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTB-0535FCS
Device Lot NumberKR988739
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received03/22/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-