• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 ACCUCATH INTRAVASCULAR CATHETER 20 GX2.25 CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 ACCUCATH INTRAVASCULAR CATHETER 20 GX2.25 CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS Back to Search Results
Model Number N/A
Device Problem Obstruction of Flow (2423)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/14/2020
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned to the manufacturer for evaluation. A lot history review (lhr) of reeu1917 showed one other similar product complaint(s) from this lot number. The complaints for this lot number (reeu1917) have been reported from the same facility.
 
Event Description
It was reported "the valve to prevent blood from coming out of the hub flips around when the needle is retracted, preventing all flow through the catheter. " it was stated this occurred twice. This report addresses the second device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameACCUCATH INTRAVASCULAR CATHETER 20 GX2.25
Type of DeviceCATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX 88780
Manufacturer Contact
kelsey erickson
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key11143484
MDR Text Key227511634
Report Number3006260740-2021-00013
Device Sequence Number1
Product Code FOZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K153298
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/08/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberN/A
Device Catalogue NumberAC0202250
Device Lot NumberREEU1917
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received12/14/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/01/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 01/08/2021 Patient Sequence Number: 1
-
-