RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Electrical /Electronic Property Problem (1198)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2020.Date of report: 08jan2021.
|
|
Event Description
|
An alarm light emitting diode (led) failure was reported.There was no patient involvement.
|
|
Manufacturer Narrative
|
H10 it was reported the power light emitting diode (led) had an illumination failure.The service technician replaced the power switch overlay and filters to address the reported problem.The unit was checked overall, run in tested, cleaned, and functionally tested.Previous investigations have determined that excessive engagement or pressure contact over the light emitting diode (led) while attempting to engage the switch due to the proximity of the led to the switch may cause a separation of the led to the encapsulant.This is subjective to the operator of the equipment which is why this is not a universal or catastrophic failure of all units.H11 g1: contact office information.
|
|