VIVONIC GMBH AQUABPLUS 2000 SUBSYSTEM, WATER PURIFICATION Back to Search Results
Catalog Number G02040107-US
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Patient Involvement (2645)
Event Date 12/01/2020
Event Type  malfunction  
Manufacturer Narrative
The plant investigation is in process. A supplemental mdr will be submitted upon completion of this activity.
 
Event Description
A user facility biomedical technician (biomed) reported that an aquabplus 2000 had a unit 1 power failure while in supply mode. While inspecting the system, the biomed found burn damage on the cable connected to stage 1 on the motor protection switch. No other damaged components were identified. There were no reports of any burning smell, smoke, sparks, or flames. To resolve the reported issue, the biomed replaced the motor protection switch along with the cables connected to it. There was no indication that the event resulted in any missed treatments or adverse events. At the time of follow up the machine was confirmed to be in service and fully operational. The replaced parts were reportedly returned to the manufacturer for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameAQUABPLUS 2000
Type of DeviceSUBSYSTEM, WATER PURIFICATION
Manufacturer (Section D)
VIVONIC GMBH
kurufuerst-eppstein-ring 4
sailauf 63877
GM 63877
Manufacturer (Section G)
VIVONIC GMBH
kurufuerst-eppstein-ring 4
sailauf 63877
GM 63877
Manufacturer Contact
matthew amaral
920 winter st
waltham, MA 02451
7816999758
MDR Report Key11143728
MDR Text Key225928120
Report Number3010850471-2021-00001
Device Sequence Number1
Product Code FIP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133829
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup
Report Date 02/10/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received01/09/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberG02040107-US
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Device AgeMO
Event Location No Information
Date Manufacturer Received02/10/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-