Model Number 16-02-80
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
There was no known patient involvement.The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).Livanova (b)(4) implemented a field safety notice for disinfection and cleaning of heater-cooler devices.The z number is z-2076/2081-2015.Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).Within the report it was stated that the device was cleaned according the instruction for use.Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report vk_20201119_13 from (b)(6) that a heater-cooler system 3t device was found to be contaminated with acid-resistant bacilli.There is no known patient involvement.
 
Manufacturer Narrative
H10: through follow-up communication livanova confirmed that the device was cleaned regularly per the instruction for use and that it was placed outside the operating theater during use.
 
Event Description
See initial report.
 
Event Description
See initial report.
 
Manufacturer Narrative
H.10: no further information is available on this specific case.Through follow-up communication performed in 2019 on previous contamination complaint for the same unit, livanova learned that the device was cleaned accordingly to the instruction for use and that it was placed outside the operating theatre.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
MDR Report Key11147473
MDR Text Key226043463
Report Number9611109-2021-00009
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 01/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/11/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-80
Was the Report Sent to FDA? No
Date Manufacturer Received03/23/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-2076/2081-2015