Model Number 16-02-80 |
Device Problem
Microbial Contamination of Device (2303)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
There was no known patient involvement.The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).Livanova (b)(4) implemented a field safety notice for disinfection and cleaning of heater-cooler devices.The z number is z-2076/2081-2015.Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).Within the report it was stated that the device was cleaned according the instruction for use.Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova (b)(4) received a report vk_20201119_13 from (b)(6) that a heater-cooler system 3t device was found to be contaminated with acid-resistant bacilli.There is no known patient involvement.
|
|
Manufacturer Narrative
|
H10: through follow-up communication livanova confirmed that the device was cleaned regularly per the instruction for use and that it was placed outside the operating theater during use.
|
|
Event Description
|
See initial report.
|
|
Event Description
|
See initial report.
|
|
Manufacturer Narrative
|
H.10: no further information is available on this specific case.Through follow-up communication performed in 2019 on previous contamination complaint for the same unit, livanova learned that the device was cleaned accordingly to the instruction for use and that it was placed outside the operating theatre.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.
|
|