Additional procode: hto.Complainant part is not expected to be returned for manufacturer review/investigation.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
It was reported that on an unknown date during an femur graft procedure, three (3) ria 2 reamer heads broke in the patient.When retrieving the bone graft, the head became disconnected from the middle tip of the long assembly.Fragments remained in the patient.The procedure was successfully completed with an unknown amount of surgical delay.The patient outcome was unknown.This report is for one (1) 10.0mm reamer head for ria 2 sterile.This is report 2 of 4 for (b)(4).
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H9: 3008812560-10/26/2020-001-c.H6: a device history record (dhr) review was conducted: part # 03.404.016s, synthes lot # 51p9256, supplier lot # 51p9256, release to warehouse date: 21 apr 2020, expiration date: 31 mar 2030, supplier: (b)(4).No ncr's were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D7b: unknown.
|