Model Number 50000000E
Device Problems Inaccurate Flow Rate (1249); Device Dislodged or Dislocated (2923)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/19/2020
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress. Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that user was getting an alert 001 (patient line open), an alert 002 (low flow) and an alert 51 (patient temperature one below control range) when user enabled normothermia mode on the arctic sun device. The user stated that the patient arrived hypothermic and wanting to warm from 28c as fast as possible on arctic sun. Hence, the user enabled normothermia mode. The water temperature was 40c, the patient temperature was 28c and flow was 0. 9lpm then stopped. Ms&s had troubleshoot low flow by disconnecting and reconnecting arctic gel pads and fluid delivery line and the flow went up to 1 then stopped. When fluid delivery line was removed, they felt little to no suction when occluding hole. Ms&s confirmed that fill tube was not connected to drain port. Ms&s helped her lower low temperature limit. Per follow up via nurse on 28dec2020, patient switched to second device and the first device sent to biomed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameARCTIC SUN 5000
Type of DeviceARCTIC SUN DEVICE
Manufacturer (Section D)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer (Section G)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key11151103
MDR Text Key226606001
Report Number1018233-2020-22436
Device Sequence Number1
Product Code DWJ
UDI-Device Identifier00801741127755
UDI-Public(01)00801741127755
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161602
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other,use
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 08/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/11/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number50000000E
Device Catalogue Number50000000E
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/04/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received08/25/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

Patient Treatment Data
Date Received: 01/11/2021 Patient Sequence Number: 1
-
-