Model Number R SERIES
Device Problems Smoking (1585); Sparking (2595)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
 
Event Description
Complainant alleged that while attempting to treat a patient (age & gender unknown) during emergency stenting, the device smoked/sparked.Complainant did not indicate that there was any adverse effect to the patient due to the reported malfunction.
 
Manufacturer Narrative
The device was returned to zoll medical united kingdom's service department.The customer's report was verified and attributed to spillage of an unknown substance on the ac charger/ac pigtail.Fluid ingress into the ac charger area can cause arcing and smoking if power is applied.However, no signs of fluid ingress were found internally.The ac charger and ac pigtail were replaced to resolve the report.The device was re-certified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R SERIES DEFIBRILLATOR
Type of Device
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
MDR Report Key11156381
MDR Text Key226314534
Report Number1220908-2021-00012
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
PMA/PMN Number
K060559/P160
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Type of Report Initial,Followup
Report Date 12/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/12/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberR SERIES
Device Catalogue NumberR SERIES
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received12/30/2020
Patient Sequence Number1
-
-