It was reported that intermittent audio output occurred.The product was evaluated.An exterior visual inspection was performed and passed.Charge and boot test was performed and passed.Manual sound test was performed and passed.Global receiver sound test was performed and passed.Receiver functional test was performed and passed.Wiggle test was performed and passed.An interior visual inspection was performed and passed.The receiver log was downloaded and reviewed finding no errors related to the complaint.The allegation was not confirmed.A probable cause could not be determined as the allegation was not found.No injury or medical intervention was reported.
|