Model Number AB14W025150150
Device Problems Burst Container or Vessel (1074); Detachment of Device or Device Component (2907)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 01/08/2021
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Physician was attempting to use a nanocross pta balloon during procedure to treat a severely calcified lesion in the mid anterior tibial artery (ata). The vessel was little tortuous. A non-medtronic inflation device used to inflate the balloon. There was no damage noted to packaging. There are no issues noted when removing device from hoop/tray. The device was prepped per ifu with no issues noted. It was reported that during balloon inflation, balloon burst/leak/rupture occurred at 8atm. All balloon fragments were not retrieved. A stent was implanted in the tibial artery to cage in the balloon fragments and maximize the lumen. No further patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameNANOCROSS 0.14 OTW PTA DILATATION CATHETER
Type of DeviceCATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer Contact
toni o'doherty
parkmore business park west
galway 
EI  
091708734
MDR Report Key11170492
MDR Text Key226809840
Report Number2183870-2021-00023
Device Sequence Number1
Product Code LIT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K082854
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial,Followup,Followup
Report Date 02/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/14/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date07/14/2021
Device Model NumberAB14W025150150
Device Catalogue NumberAB14W025150150
Device Lot NumberA675828
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received02/11/2021
Was Device Evaluated by Manufacturer? No Answer Provided
Date Device Manufactured07/15/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 01/14/2021 Patient Sequence Number: 1
-
-