It was reported that during a cryo ablation procedure, the balloon catheter was difficult to pass through the sheath.During aspiration, there was more air aspirated than expected, indicating that air ingress occurred.It was suspected that an air embolism or clot had formed, and the patient subsequently experienced hemiparesis on the right side.The case was completed with cryo.After the case and magnetic resonance imaging (mri) showed that the patient had experienced a stroke.The patient's hospitalization was extended for recovery.It was also reported that the data files could not be downloaded from the console.No further patient complications have been reported as a result of this event.
|
Product event summary: the data files were returned and analyzed.The files did not show any system notices on the date of the event.Clinical issues (a stroke, air embolism, thrombus, and hemiparesis) were encountered post procedure.In conclusion, the physical product was not returned for investigation.There is no indication of a malfunction of the sheath related to the adverse event.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|