C.R. BARD, INC. (BASD) -3006260740 PER-Q-CATH PLUS 2F SL WITH STIFFENING STYLET; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
|
Back to Search Results |
|
Model Number 4132105 |
Device Problems
Leak/Splash (1354); Material Perforation (2205); Device Damaged by Another Device (2915)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/21/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that at the time of passage, the guidewire perforated the catheter, causing a leak.The patient was submitted to a venous dissection.No other information was provided.
|
|
Manufacturer Narrative
|
The information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The device has not been returned to the manufacturer for evaluation, as the device was discarded after the event occurred.A lot history review (lhr) of rect0098 showed two other similar product complaint(s) from this lot number.Device not returned for evaluation.
|
|
Search Alerts/Recalls
|
|
|