Device was inspected.Evaluation determined that the reported issue was not confirmed, unit showed an image , however chipped from ¿a-rubber¿ glue was observed.Based on evaluation findings the reported issue was not confirmed.Physical defects found likely attributed to handling issue.Root cause of the issue is unknown as the reported failure cannot be duplicated.
|
This report is being supplemented to provide additional information based on the legal manufacturer¿s final investigation.As a result of the device investigation, olympus was not able to conclusively determine the cause of the reported event, as it was not duplicated during testing.It is possible that the user may have had dirty contacts or a poor connection, resulting in no image being displayed.Based on the legal manufacturer¿s investigation, the device history record (dhr) was reviewed and there were no problems found during the manufacturing of the device that would cause or contribute to the reported issue.The device met all specifications at the time of shipment.
|