• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TANDEM DIABETES CARE T:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY; AUTOMATED INSULIN DOSING, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TANDEM DIABETES CARE T:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY; AUTOMATED INSULIN DOSING, THRESHOLD SUSPEND Back to Search Results
Model Number 1000096
Device Problem Leak/Splash (1354)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/13/2020
Event Type  malfunction  
Manufacturer Narrative
No product was returned for evaluation.Should new relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that multiple cartridges were leaking at the septum.Customer loaded a new cartridge to address the issue.There was no adverse impact to the customer's blood glucose level.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY
Type of Device
AUTOMATED INSULIN DOSING, THRESHOLD SUSPEND
Manufacturer (Section D)
TANDEM DIABETES CARE
11075 roselle street
san diego CA 92121
Manufacturer Contact
mick trier
san diego, CA 92121
8584011451
MDR Report Key11186493
MDR Text Key229933319
Report Number3013756811-2021-10010
Device Sequence Number1
Product Code OZO
UDI-Device Identifier00853052007264
UDI-Public00853052007264
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P180008
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 01/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/18/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number1000096
Device Catalogue Number1000886
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Manufacturer Received01/13/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/01/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age38 YR
-
-