Model Number 72200616
Device Problems Mechanical Jam (2983); Intermittent Energy Output (4025)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/11/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4). The reported device, used in treatment, was received for evaluation. There was a relationship found between the returned device and the reported incident. A visual inspection was performed on the product and observed no issues. A functional evaluation revealed a jammed motor/ blade stall. A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended. A complaint history review concluded this was a repeat issue. The complaint was confirmed and the root cause has been associated with a mechanical component failure. Factors that could have contributed to the reported event include corrosion in the gearbox/motor assembly from cleaning and sterilization methods and the chemicals involved over a period of time or one or more of the motor phases shorting out. No containment or corrective actions are recommended at this time.
 
Event Description
It was reported that the motor drive unit stops and goes during a knee procedure. The malfunction was solved with no delays or patient harm using a back up device. Results of investigation have shown the shaver had a jammed motor/blade stall error which makes it a reportable event. All available information has been disclosed. If additional information should become available, a supplemental report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMOTOR DRIVE UNIT HAND CNTRL PWRMX EL
Type of DeviceSAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key11190242
MDR Text Key227591889
Report Number1643264-2021-00203
Device Sequence Number1
Product Code HAB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/18/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received01/18/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number72200616
Device Catalogue Number72200616
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/28/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/13/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage

-
-