DATEX-OHMEDA, INC. AISYS; ANESTHESIA GAS MACHINE Back to Search Results
Device Problem No Display/Image (1183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/23/2020
Event Type  malfunction  
Manufacturer Narrative
Customer contact reports no patient information available.A ge healthcare service representative performed a checkout of the system and confirmed the reported issue.The display monitor was replaced by the customer to repair the issue.
 
Event Description
The hospital reported that the display went blank and mechanical ventilation stopped during a case.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AISYS
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer Contact
john szalinski
3030 ohmeda drive
madison, WI 
MDR Report Key11193749
MDR Text Key227579409
Report Number2112667-2021-00178
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172045
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 01/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/19/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/23/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/23/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-